CriticalCritical vulnerability
CVE-2026-105279 (CVSS 9.3)
NVD · officialPublished Oct 16, 2026Risk 50/100
In the default configuration, the archive accepts DICOM associations from an unregistered calling AE. A remote peer can use C-FIND and C-GET to enumerate and retrieve patient information and medical images. It can also use C-STORE to file fabricated study content under an existing patient identity.
Technical details
CVSS
9.3
AV:NetworkAC:LowPR:NoneUI:None
Evidence and sources
This record is attributed to NVD. Exploitation status and remediation guidance are kept separate from the vulnerability's technical severity.
Open primary source